You often use orthogonal regression in clinical chemistry or a laboratory to determine whether two instruments or methods provide comparable measurements. If the confidence interval for the constant term contains zero and the interval for the linear term contains 1, then you can usually conclude that the measurements from the two instruments are comparable.
You should also examine the plot with the fitted line to determine how well the model fits the data.
Predictor | Coef | SE Coef | Z | P | Approx 95% CI |
---|---|---|---|---|---|
Constant | 0.64441 | 1.74470 | 0.3694 | 0.712 | (-2.77513, 4.06395) |
Current | 0.99542 | 0.01415 | 70.3461 | 0.000 | (0.96769, 1.02315) |
Variable | Variance |
---|---|
New | 1.07856 |
Current | 1.19840 |
In these results, the confidence interval for the constant term is approximately (−3, 4). Because the interval contains 0, this part of the analysis does not provide evidence that the measurements from the two instruments differ.
The confidence interval for the linear term is approximately (0.97, 1.02). Because the interval contains 1, this part of the analysis does not provide evidence that the measurements from the two instruments differ.
Because neither interval provides evidence that the measurements from the two instruments differ, you usually conclude that the measurements are comparable. You should also verify that the model fits the data well by examining the plot with the fitted line and the residual plots.
This plot shows an example of measurements from two instruments or methods that are comparable. The points follow the fitted line with minimal scatter and without any pattern that reveals systematic differences between the methods.
In the results below, the confidence intervals for the coefficients do not provide evidence that the measurements of the two instruments differ. However, the plot shows that points do not fall close to the line, which indicates that the measurements from the two instruments are not comparable. Because the data do not fit the equation, the usual conclusion is that the instruments differ.
Predictor | Coef | SE Coef | Z | P | Approx 95% CI |
---|---|---|---|---|---|
Constant | -0.00000 | 0.215424 | -0.0000 | 1.000 | (-0.422224, 0.42222) |
New | 1.00000 | 0.517586 | 1.9320 | 0.053 | (-0.014450, 2.01445) |