Statistics for VDA 5 in Nested Gage R&R Study

The following equations give the uncertainties for the component.
Resolution (uRE)
Repeatability of measured object (uEVO)
√(Variance component for repeatability)
Reproducibiltiy of operators (uAV)
√(Variance component for operator)
Other factors (uREST)
Measurement Process (uMP)

% of Total

The percentage of the measurement process from each component has the following equation:

%QMP

The calculation for the percent of the tolerance uses the uncertainty due to the measurement process.

Notation

TermDescription
nthe number of measurements
xithe measurement of the ith part
the mean of n measurements
REThe resolution of the measurement process. The calculations consider the uncertainty due to resolution if the specifications for the analysis include the resolution.
xmthe reference measurement
FiThe uncertainty due to additional factors. The values are from specifications for the analysis.
athe number of other factors that contribute to uncertainty in the specifications for the analysis
uCALThe uncertainty due to calibration. The value is 0 unless the specifications for the analysis include the value.
uEVRThe uncertainty due to repeatability at a reference value. The value is 0 unless the specifications for the analysis include the value.
uBIThe uncertainty due to bias. The value is 0 unless the specifications for the analysis include the value.
uLINThe uncertatinty due to linearity. The value is 0 unless the specifications for the analysis include the value.
uia single component of uncertainty
k1

Minitab uses the default value of 6 standard deviations from a standard normal distribution to represent 99.73% of your measurements.

To change this value, see the Options subdialog box. For example, use the multiplier 5.15 to represent 99% of your measurements.